12.
 
CALIBRATION 
 
The  GBK/GFK  scales can  be  calibrated  using  kilogram  weights or  using 
pounds weights, depending on the weighing unit selected at the time of 
calibration. 
 
 
To  start  the  calibration,  either  get  into  the  calibration  section  through 
the Scale Settings (“FUnC 3”- see Section 13.3) or turn the scale off and 
switch  on  again  then  press  [Tare]  during  the  self-test.    Enter  code 
number  0000  and  press  [Tare].  This  will  take  you  directly  to  the 
calibration section. 
 
 
The display will show "UnLoAd"  
 
 
Remove  all  weight  from  the  platform  and  then  press  the  [Tare]  key 
when  the  display  is  stable.  After  the  Zero  point  is  set,  the  display  will 
show “Ld  xx”. Place the suggested calibration mass on the platform. It is 
best to use a weight close to the full capacity of the scale.  If the mass is 
different from the displayed value, enter the value of the mass in whole 
numbers. The kg or the lb symbol will be on to show the active unit. 
 
 
Press the [Tare] key when the stable indicator is on.   
 
 
The  scale  will  calibrate  to  the  mass.  When  complete,  it  will  display 
“PASS”  and  then  either  display  “S8    CAL”  (if  entered  the  calibration 
section  through  the  Scale  Settings  as  per  section  13.3)  or  return  to 
normal weighing (if entered directly). Remove the calibration mass. 
 
 
If an error message “FAIL H” or “FAIL L” is shown, re-check the calibration 
and repeat. If the error cannot be corrected contact your supplier.