Scale Calibration Procedure (lbs. or kg) with Remote Display 
 
7.  Make sure there is no load on the scale platter. 
 
8.  Calibrate a zero load.  Press the right arrow (►) button once when the remote display reads CAL 1 and there is no load 
on the scale platter. 
 
 
 
Figure 44. 
 
 
 
9.  Place a half load (15.00 lbs. or 7.500 kg) on the center of the scale platter then press the right arrow (►) button. 
 
 
 
Figure 45.
 
 
 
10.  Add an additional half load (15.00 lbs. or 7.500 kg) to the existing half to simulate a full load, center entire load then press 
the right arrow (►) button. 
 
 
 
Figure 46.
 
 
 
11.  Remove half of the load (15.00 lbs. or 7.500 kg), center the remaining load then press the right arrow (►) button.  
 
 
 
Figure 47.