Graham Field 240 Manual Do Utilizador

Página de 12
9
Measurement Procedures
1.  To fasten the cuff, press the hook material 
firmly against the fuzzy pile material. Make 
sure  the  stethoscope  chestpiece  sits  over 
the brachial artery. If necessary, readjust the 
cuff.
2.  Place your arm on a table so that the cuff is at 
the same level as your heart.
CAUTION:  Before  you  put  the  stethoscope  binaural  in 
your ears, make sure the eartips are securely fastened to 
the binaural.
3.  Insert  eartips  into  your  ears. Adjust  the  metal 
binaural for comfort.
4.  Clip the gauge to a hard cover book on the table 
in front of you.  Hold the inflation bulb in your 
right hand.  Close the air release valve attached 
to the inflation bulb by turning it clockwise.
5.  To inflate the cuff, slowly and steadily pump the 
inflation bulb.
Caution:  Do  not  inflate  the  cuff  above  280  mmHg. You  may  damage  the 
instrument or injure yourself. If you are unsure, consult your doctor for the 
correct inflation level.
Note:  If  you  KNOW  your  SYSTOLIC  reading,  pump  the  inflation  bulb  until 
the gauge reaches approximately 50 mmHg ABOVE your average SYSTOLIC 
reading. If you DO NOT KNOW your SYSTOLIC reading, pump the inflation 
bulb until the gauge needle reaches 180 mmHg.
6.  Slowly open the air release valve by turning it counter-clockwise.  
Release the air at a rate of 2-3 mmHg per second.