National Instruments SCC-TC01 Manual Do Utilizador

Página de 20
© National Instruments Corporation
15
SCC-TC Series Thermocouple Input Modules User Guide
Cold-Junction Sensor
Cold-junction sensor accuracy
............... ±0.4 °C max from 15 to 35 °C,
±0.75 °C max from 0 to 15 °C 
and 35 to 55
 °C
Output..................................................... 1.91 V (0 
°C) to 0.58 V (55 °C)
Note
The accuracy specification includes the combined effects of the temperature sensor 
accuracy and the temperature difference between the temperature sensor and any 
thermocouple connector. The temperature sensor accuracy includes component tolerances, 
temperature drifts, and self-heating effects. It does not include measurement device errors.
Open Thermocouple Detection
Pull-up resistor ....................................... 10 M
Ω
Bias resistor............................................ 10 M
Ω
Maximum field wire gauge .................... 16 AWG
Power Requirement
Analog power ......................................... 60 mW max
+15 V .............................................. 2 mA max
–15 V............................................... 2 mA max
Digital power (
+5 V).............................. 0.0 mW
K
–200 to –100
–100 to 600
600 to 1,200
2.5
1.25
2
1.25
0.67
0.45
N
–200 to –100
–100 to 1,300
2.5
2
2
0.7
R
–50 to 0
0 to 100
100 to1,600
3
2
1.75
1.24
0.82
0.4
S
–50 to 0
0 to 1,400
1,400 to 1,600
3
2
2.5
1.3
1
1
T
–200 to –100
–100 to 400
2.5
1
2
0.9
Thermocouple 
Type
Temperature 
Range (
°C)
Maximum 
Inaccuracy (
±°C)
Typical 
Inaccuracy (
±°C)