Fluke 1744 Manual Do Utilizador

Página de 73
1744/1743 
Users Manual 
52 
General Specifications 
RMS Logging Slow Voltage Variations 
Logging values: Mean value 
RMS values averaged over interval length 
Min, Max values 
Averaging with selectable averaging time from 
0.5 periods to 5 s 
Max value 
Max 10 ms RMS value per interval 
Min value: 
Min 10 ms RMS value per interval 
Current Logging Values 
Mean value 
RMS values averaged over interval length 
Max value 
Highest RMS value per interval 
Events Dips, Swells, Interruptions 
Limit value 
Variable 
Lower limit: 0 to 95% V
Upper limit: 105 to 120% V
N
 
Set in PQ Log 
Range 
0 to V
I
 + 20% 
Logging value 
½ period RMS value 
Operating uncertainty 
< 2% of V
I
 
Response time 
½ line power period 
Flicker 
Logging value 
Flicker severity (Pst / Plt) according to 
IEC 61000-4-15 
Intrinsic uncertainty Pst 
< 5% of m. v 
Measuring range Pst 
0.4 to 4