9011 Dual-Well Calibrator
User’s Guide
42
Command Description
Command
Format
Command
Example
Returned
Returned
Example
Acceptable
Values
Set ALPHA calibration parameter
to
n
al[pha]=n
al=0.0038433
0.0032 to
0.0042
Read DELTA calibration parameter de[lta]
de
de: 9.99999
de: 1.46126
Set DELTA calibration parameter
to
n
de[lta]=
n
de=1.45
1.0 to 1.9
Read BETA calibration parameter
(cold side only)
be[ta]
be
be: 9.999
be: 0.342
Set BETA calibration parameter to
n (cold side only)
be[ta]=
n
be=0.342
–25.0 to 25.0
Miscellaneous (not on menus)
Read firmware version number
*ver[sion]
*ver
ver.9999,9.99
ver.9011,1.11
Read structure of all commands
h[elp]
h
list of commands
Read all parameters
all
all
list of parameters
and their values
Legend:
[] Optional Command data
/ Alternate characters or data
{} Returns either information
n Numeric data supplied by user—may be entered in decimal or exponential notation
9 Numeric data returned to user
x Character data returned to user
Note:
When DUPLEX is set to FULL and a command is sent to READ, the command is returned followed by
a carriage return and linefeed. Then the value is returned as indicated in the RETURNED column.
Note:
Commands that are preceded by a “C:” or an “H:” access the cold or hot side respectively. If the “C:” or
“H:” are not used, the command accesses the cold side.
Controller Communications Commands continued