eutech-instruments cyberscan con 11 conductivitytds meter Manual Do Utilizador
Instruction Manual
CON 11/110
43
8.4
P3.0 Viewing Electrode Data
Program 3.0 has five “view only” options that let you check the electrode’s
parameters for diagnostic purposes. These options show you the effective cell
constant for each range. The cell constant is adjusted
according to your calibration.
From measurement mode
parameters for diagnostic purposes. These options show you the effective cell
constant for each range. The cell constant is adjusted
according to your calibration.
From measurement mode
1. Press the SETUP key to enter Set Up menu.
2. Press the MI/▲ or MR/▼ keys until you view
parameter P3.0.
3. Press
the
HOLD/ENTER
key repeatedly to view the
effective cell constant for each range.
4. When you have scrolled through all the electrode
data, you will automatically return to the subgroup
menu.
menu.
5. Press the CAL/MEAS key to return to measurement
mode.
NOTES: Cell constants will degrade with time and usage.
You can use this feature to alert you to the need for a new
electrode prior to total failure.
You can use this feature to alert you to the need for a new
electrode prior to total failure.
SETUP
1.00
p3.1
SETUP
p 3.0
ele
SETUP
p 3.2
1.00
ENTER
ENTER
K=
K=
Figure 36: P3.0-Viewing
electrode data for each
measurement range