Instruction Manual 
 
Cyanide Electrode
 
 
17
Low Slope or         
standards contaminated     
 
prepare fresh standards      
No Slope               
or incorrectly made 
                      
 
 
 
ISA not used              
 
use recommended ISA 
 
                         
standard used as ISA    
 
use ISA 
 
                         
electrode exposed        
 
check section entitled 
                         
to complexing agents      
 
Complexation 
 
                         
air bubble on membrane          
 
remove bubble by 
                             
 
 
 
 
 
re-dipping probe 
 
 
"Incorrect Answer"    incorrect scaling of     
 
plot millivolts on the linear 
(but calibration       
semi-log paper       
 
 
axis.  On the log axis, be sure 
curve is good)                                
 
 
 
concentration numbers within 
        each 
decade 
are 
increasing 
        easing 
concentration. 
 
                         
incorrect sign          
 
 
be sure to note sign 
                                                
 
 
 
of millivolt number correctly 
 
                         
incorrect standards       
 
prepare fresh standards 
 
                         
wrong units used          
 
apply correct conversion factor: 
 
 
 
 
 
                                   10
-3
M = 26.0 ppm CN- 
                                                   
                         
complexing agents       
 
check section entitled 
                        
in sample                    
 
Complexation