Calcium Electrode 
 
Instruction Manual 
 
18
                                                                               
 
between samples 
 
 
Drift (reading       
samples and standards    
 
allow solutions to come to room 
slowly changing      
at different             
 
 
temperature before measurement 
in one direction)    
temperatures               
 
 
 
                       
electrode exposed         
 
soak electrode in 
                       
to interferences           
 
calcium standard 
 
                       
incorrect reference       
 
use recommended 
                       
filling solution           
 
filling solution 
 
                       
incorrect pH               
 
adjust to pH 3-10 
                                               
 
 
 
with NaOH or HCl 
 
Low Slope or         
standards contaminated      
 
prepare fresh standards 
No Slope             
or incorrectly made 
 
                       
defective electrode       
 
check electrode operation 
 
                                            
                       
air bubble on membrane   
 
remove bubble by 
                                               
 
 
 
re-dipping probe 
 
                       
electrode exposed to     
 
soak electrode in 
                    
 
interferences              
 
calcium standard 
 
                       
standard used as ISA     
 
use ISA 
 
 
 
 
ISA not used               
 
use ISA