To operate the system: 
1) 
Connect J26 connector of Wave Plus by cable (analog output) to the analog A/D interface of 
the external device; 
2) 
If  required,  connect  the  trigger  cable  to  the  external  device.  Check  trigger  electric  features 
(voltage and timing) adopted by the system (see “Technical specifications” chapter); 
3) 
Connect Wave Plus power supply and switch on the system. 
 
The system can be powered only with Friwo FW7555M/09 power supply provided by the 
producer;  the  use  of  other power  supply  units  can  cause  electric  shock  and  damage  the 
system. 
 
If used in analog mode, the system  does not require a Personal Computer; after switching on the base 
unit, EMG and basographic signals are available in continuous on analog output connector J26. 
 
 
4.1.1  Supported external devices 
 
The Wave Plus system can be connected to analog devices commonly used in clinical and scientific en-
vironments; these are the features required by systems: 
 
  At least 8 or 16 EMG data collection channels, configured for signals with Zout = 100 ohm  and 
an amplitude of ± 2,5 V; 
  At  least  two  analog  channels  to  acquire  footswitches,  configured  for  signals  with  Zout  =  100 
ohm and an amplitude of + 2,5 V; 
  If  also  accelerometer  signals  has  to  be  transmitted  to  the  analog  device,  another24  or  48  data 
collection channels (3 for each tri-axis accelerometer) are needed; 
  It complies with IEC 60950. 
 
 
The use of external devices non-complying with IEC 60950 can cause electric shock and 
damage  the  system.  The  external  devices  connected  to  Wave  Plus  with  an  analog 
interface, do not have any control over the system acquisition or settings. Any third party 
system  used  with  Wave  Plus  has  to  be  placed  out  of  the  patient  area,  and  must  be 
powered via an isolation transformer or power supply conform to the IEC 60601-1.