Cometa Sas di A. Dellacorna & C. WPTX Manual Do Utilizador
WAVE PLUS - USER MANUAL
C
OD
.
WPMEMG
REV
01
02/11/2011
D
ATA PRIMA EMISSIONE
Pag. 16 di 37
TRIG_IN
STROBE I/O
GND
1
2
3
4
4
J22
TRIG_OUT
4.2.1.1 Trig in
Data capture can be synchronised with an external trigger signal. If required, connect the trigger cable to
the external device after checking that the trigger control complies with the electric features required by
the system (see “Technical Specifications”).
the external device after checking that the trigger control complies with the electric features required by
the system (see “Technical Specifications”).
Trigger logic:
1)
Data acquisition if the “trigger in” signal is at “1” logic level
2)
No data acquisition if the “trigger in” signal is at “0” logic level
The default condition of the line is “1”.
Storing data on the hard disk of your computer is managed by the acquisition and processing SW; for
information about the data file format and processing, refer to the user manual of the SW.
information about the data file format and processing, refer to the user manual of the SW.
4.2.1.2 Trig out
A “trigger out” channel shows the system status:
1) “Trig out” signal at “1” logic level when the system is acquiring
2) “Trig out” signal at “0” logic level when the system is not acquiring
4.2.1.3 Strobe IO
The line "Strobe IO" is used for synchronization of a second Wave Plus receiver for the expansion to
32channelsof the system. This line is used by the system and should only be connected to another Wave
Plus unit, according to information provided by Cometa.
32channelsof the system. This line is used by the system and should only be connected to another Wave
Plus unit, according to information provided by Cometa.