Cometa Sas di A. Dellacorna & C. WPTX Manual Do Utilizador

Página de 37
 
WAVE PLUS - USER MANUAL 
C
OD
.
 
WPMEMG
 
REV 
01 
02/11/2011
 
D
ATA PRIMA EMISSIONE
 
 
Pag. 16 di 37 
 
TRIG_IN
STROBE I/O
GND
1
2
3
4
J22
TRIG_OUT
 
 
 
 
4.2.1.1  Trig in 
 
Data capture can be synchronised with an external trigger signal. If required, connect the trigger cable to 
the external device after checking that the trigger control complies with the electric features required by 
the system (see “Technical Specifications”).  
 
Trigger logic: 
 
1) 
Data acquisition if the “trigger in” signal is at “1” logic level  
2) 
No data acquisition if the “trigger in” signal is at “0” logic level  
 
The default condition of the line is “1”. 
Storing data on the hard disk of your computer is managed by the acquisition and processing SW; for 
information about the data file format and processing, refer to the user manual of the SW. 
 
4.2.1.2  Trig out 
 
A “trigger out” channel shows the system status: 
 
1)  “Trig out” signal at “1” logic level when the system is acquiring  
2)  “Trig out” signal at “0” logic level when the system is not acquiring  
4.2.1.3  Strobe IO 
 
The  line  "Strobe  IO"  is  used  for  synchronization  of  a  second  Wave  Plus  receiver  for  the  expansion  to 
32channelsof the system. This line is used by the system and should only be connected to another Wave 
Plus unit, according to information provided by Cometa.