Insulet Corporation 019 Manual Do Utilizador

Página de 190
124
10
BG Meter Error 4.
Strip fill problem.
Retest with a new 
strip. 
Not enough blood or 
control solution was 
applied or more was 
added after the meter 
began to count down.
• The test strip may have 
been damaged or moved 
during testing.
• The sample was improp-
erly applied.
There is a problem with the 
meter.
1. Retest using a new test strip. Do not press any buttons or Soft Keys 
while applying blood or control solution to the test strip.
2. If the error persists, call Customer Care or perform another blood test.
BG Meter Error 5.
Strip problem.
Retest with a new 
strip. 
The meter has detected a 
problem with the test 
strip, possible cause is 
test strip damage.
1. Retest using a new test strip. Do not press any buttons or Soft Keys 
while applying blood or control solution to the test strip.
2. If the error persists, call Customer Care.