BG Meter Error 4.
Strip fill problem.
Retest with a new
strip.
strip.
•
Not enough blood or
control solution was
applied or more was
added after the meter
began to count down.
control solution was
applied or more was
added after the meter
began to count down.
• The test strip may have
been damaged or moved
during testing.
during testing.
• The sample was improp-
erly applied.
•
There is a problem with the
meter.
1. Retest using a new test strip. Do not press any buttons or Soft Keys
while applying blood or control solution to the test strip.
2. If the error persists, call Customer Care or perform another blood test.
BG Meter Error 5.
Strip problem.
Retest with a new
strip.
strip.
•
The meter has detected a
problem with the test
strip, possible cause is
test strip damage.
strip, possible cause is
test strip damage.
1. Retest using a new test strip. Do not press any buttons or Soft Keys
while applying blood or control solution to the test strip.
2. If the error persists, call Customer Care.
Alert error message
Possible causes
Action