 INSTALLATION
4.7. Balancing the double pantograph arm
Figure  M
If the double pantograph arm does not stay in a stable position, adjust the spring tension by using an 8mm Allen wrench about 20cm long. 
• To adjust the arm (a) connected to the extension: position it as shown in the fi gure and place the wrench at point A.
 NOTE:
TIGHTEN (clockwise) if the arm tends to move down.
LOOSEN (anti-clockwise) if the arm tends to move up.
• To adjust the arm (b) connected to the generator:  put the 2 covers (c) and (d) on the front pivot point, move the arm (b) into a horizontal position 
and insert the key in point B.
 NOTE:
TIGHTEN (clockwise) if the arm tends to move down.
LOOSEN (anti-clockwise) if the arm tends to move up.
4.8. Adjusting the double pantograph arm end-stops
Figure  N
If the end-stops of the double pantograph arm need to be adjusted, work on the screws (- l n) shown in the diagram. 
 NOTE:
To adjust the screw (n) on the articulated joint of the extension side place the relevant arm (a) in a horizontal position.