Vieworks Co. Ltd. FXRD-1012NAW Manual Do Utilizador

Página de 151
 
 
Rev.1.0
 
Page 101 of 151 
VW40-152-009
 
VIVIX-S 1012N User Manual 
 
 
You can only change the displayed direction of an image from Detector Direction 
Compensation
. The direction of an original image cannot be changed. 
 
 
 
Effective Area can be changed within the effective range which has been configured at 
the time of the first shipment of detector. 
 
 
 
Once Use offset refresh is used, the offset refresh is processed automatically as follows. 
 
Checks temperature difference of the detector between current temperature and the 
previous one at the time of offset refresh in every setting time from Time Interval
 
If the temperature difference is more than the one set in Temperature Interval, the 
offset refresh is carried out a number of times set in Number of shot
 
If you use VXvue, a pop-up message will be displayed before the offset refresh. 
 
Auto offset refresh will be performed by clicking OK button. 
 
 
 
OSF is the auxillary function for stabilizing pixel value of an image to the default value 
rapidly. 
 
The pixel value can be displayed equally by using OSF under the environment where the 
pixel value of X-ray image is required to display. 
 
Whether to use OSF or not does not affect the X-ray image quality.