Xplore Technologies Corp. EM75 Manual Do Utilizador

Página de 41
19 
 
Table 0-3: Guidance and manufacturer’s declaration – electromagnetic immunity – for ME equipment and ME systems that are 
not Life-Supporting 
The XSlate L10 Rugged Tablet PC is intended for use in the electromagnetic environment specified 
below. The customer or the user of the XSlate L10 Rugged Tablet PC should assure that it is used in 
such an environment. 
 
Immunity test 
Test level 
Compliance level 
Electromagnetic environment – guidance 
 
 
 
Portable and mobile RF communications equipment should be 
used no closer to any part of the XSlate L10 Rugged Tablet PC 
including cables, than the recommended separation distance 
calculated from the equation applicable to the frequency of the 
transmitter. 
 
Recommended separation distance 
Conducted RF 
(IEC 61000-4-6) 
3 Vrms 
150 kHz to 80 MHz
 
3 Vrms 
150 kHz to 80 MHz
 
𝑑 = [
3.5
𝑉
1
] √𝑃 
Radiated RF 
(IEC 61000-4-3) 
3 V/m 
80 MHz to 2.5 GHz 
3 V/m 
80 MHz to 2.5 GHz 
𝑑 = [
3.5
𝐸
1
] √𝑃 80 𝑀𝐻𝑧 𝑡𝑜 800 𝑀𝐻𝑧 
𝑑 = [
7
𝐸
1
] √𝑃 800 𝑀𝐻𝑧 𝑡𝑜 2.5 𝐺𝐻𝑧 
 
 
 
where is the maximum output power rating of the transmitter 
in watts (W) according to the transmitter manufacturer and is 
the recommended separation distance in metres (m). 
 
Field strengths from fixed RF transmitters, as determined by an 
electromagnetic site survey, 
(a)
 should be less than the 
compliance level in each frequency range. 
(b)
 
Interference may occur in the vicinity of equipment marked with 
the following symbol: 
 
 
Notes: 
NOTE 1    At 80 MHz and 800 MHz, the higher frequency range applies. 
NOTE 2    These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption 
and reflection from structures, objects and people. 
 
(a) 
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land 
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically 
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site 
survey should be considered. If the measured field strength in the location in which the XSlate L10 Rugged 
Tablet PC is used exceeds the applicable RF compliance level above, the XSlate L10 Rugged Tablet PC should be 
observed to verify normal operation. If abnormal performance is observed, additional measures may be 
necessary, such as reorienting or relocating the XSlate L10 Rugged Tablet PC. 
(b)
 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.