Patient Remote Control
45
do not have any, the link is completed immediately and you will see 
this message: 
However, if programs are present in either device during linking, you 
will be required to respond to one or more “decision” screens to guide 
the Remote Control to complete the linking with the desired program 
set saved to the desired device (Remote Control or Stimulator). You 
may also need to enter the clinician’s password.
If an error occurs during the process or if the password is incorrect 
linking will be aborted. For further information, including informa-
tion about the clinician’s password, see “To Clear Link” following.
Searching
In the event of communication interference, the Remote Control will 
automatically begin “searching” for the Stimulator. Try to reposition