                      Sensor Headset Model 603-B Technical Manual 
Guidance and manufacturer’s declaration – electromagnetic immunity 
The Model 603-B is intended for use in the electromagnetic environment specified below.  The customer or the 
user of the Model 603-B should assure that it is used in such an environment. 
Immunity Test 
IEC 60601 test level 
Compliance 
level 
Electromagnetic environment - guidance 
 
 
 
 
 
 
 
 
 
Conducted RF 
IEC 61000-4-6 
 
 
Radiated RF 
IEC 61000-4-3 
 
 
 
 
 
 
 
 
 
3 Vrms 
150 kHz to 80 MHz 
 
 
3 V/m 
80 MHz to 2.5 GHz 
 
 
 
 
 
 
 
 
 
3 Vrms 
 
 
 
3 V/m 
Portable and mobile RF communications 
equipment should be used no closer to any part of 
the Model 603-B, including cables, than the 
recommended separation distance calculated from 
the equation applicable to the frequency of the 
transmitter. 
 
Recommended separation distance 
 
 d = 1.2
P
 
 
 
d = 1.2
P
   
80 MHz to 800 MHz
 
d = 2.3
P
   
800 MHz to 2.5 GHz 
 
 
where P is the maximum output power rating of 
the transmitter in watts (W) according to the 
transmitter manufacturer and d is the 
recommended separation distance in meters (m). 
 
Field strengths from fixed RF transmitters, as 
determined by an electromagnetic site survey,
should be less than the compliance level in each 
frequency range.
b
 
 
Interference may occur in the vicinity of 
equipment marked with the following symbol: 
 
 
 
NOTE 1    At 80 MHz, the higher frequency range applies. 
NOTE 2    These guidelines may not apply in all situations.  Electromagnetic propagation is affected by 
absorption and reflection from structures, objects and people. 
a
    Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land 
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically 
with accuracy.  To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site 
survey should be considered.  If the measured field strength in the location in which the Model 603-B is used 
exceeds the applicable RF compliance level above, the Model 603-B should be observed to verify normal 
operation.  If abnormal performance is observed, additional measures may be necessary, such as re-orienting or 
relocating the Model 603-B 
b
    Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. 
 
 
 
 
Table 206 