Rayence Co. Ltd. RY1417W Manual Do Utilizador

Página de 83
52 
 
R-USM-020 
 
 
 
 
Bright Calibration Point 
To  gain  correction,  bright  frame  and  dark  frame  should  be 
acquired.  The  dark  frame  is  needed  only  one  frame.  The  bright 
frame  is  recommended  to  be  acquired  more  than  2  different 
levels  of  median  values  of  bright  frames.  The  X-ray  source 
condition  will  be  recommended  the  tube  energy  level  of  70kVp 
with variable tube current and exposure time. To acquire at least 
2  frames  at  same  condition  will  be  recommended.  The  median 
values of bright frames are below.  
(Additional 21mm Al filter is recommended for calibration) 
 
 
 
The Purpose of Bright Calibration 
The center of the non calibrated image is brighter than the edge 
due to hill effect of X-ray exposure. Generally, the intensity of X-
ray  flux  at  center  region  of  exposed  area  is  higher  than 
surroundings  due  to  the  X-ray  expose  like  cone  shape.  A 
calibration  process  is  used  to  compensate  for  this  effect. 
Generally, called it ‘Flat Field Correction’(Bright calibration). 
 
 
Point 
Median Value [LSB 
500 ~ 
1000 
2900 ~ 
3400 
6200~ 
6700 
9500 ~ 
10000 
Table 5 : Median value 
                The  
                calibration  
                range of 
bright is can be 
select by which 
exposure level is 
maximum level that 
user want to use. If 
the maximum level 
of user want to use 
is ‘6500’ in this case 
the level is 
contained in Bright 
point of ‘3’(refer 
‘Table 5 : Median 
value’). The meaning 
is you don’t have to 
make
 
bright point 
for ‘4’(In this case, it 
will be does not 
working  if  you get 
image on higher 
level than maximum 
bright point.)