Data Critical Corporation AVTS1010 Manual Do Utilizador

Página de 55
AlarmView™ System User’s Guide 
 
 
 
{insert part #} Page 33 
 
 
4. Tap to select desired time interval from pull-down menu. 
 
5. Aim programmer infrared port directly at infrared port on 
front of transmitter and tap SEND to program configuration 
changes into transmitter. A “Programming OK” message will 
be sent to pager(s) assigned to transmitter. 
 
6. Repeat programming for every transmitter in system. 
 
7. Tap Exit button to return to SysMgr. 
 
 
Section 5: Testing and Periodic 
Maintenance 
 
All tests described in this section can be performed without 
disassembling the units in the AlarmView System. Tests should be 
conducted prior to initial installation, when adding a new unit to the 
system, and before returning equipment to service after repairs or routine 
maintenance. 
 
!
!
!
 
 
 
If any component of the system fails to perform as specified, do not 
return it to use. Contact Data Critical Corp. regarding repair or 
replacement of the equipment. 
 
Testing System Safety 
 
AlarmView safety tests meet the standards of, and are to be performed 
in accordance with, CAN/CSA-C22.2 No. 601.1 and UL 2601-1, Medical 
Electrical Equipment – Part 1: General Requirements for Safety, for 
instruments classified as Class 1 and Not Classified. 
 
!
!
!
!
!
!
 
 
 
 
 
 
Equipment Needed is a Safety Analyzer that must meet , 
CAN/CSA-C22.2 No. 601.1 and UL 2601-1 specifications. Bio-
Tek Instruments, Inc. or compatible.