Data Critical Corporation AVTS1010 Manual Do Utilizador
AlarmView™ System User’s Guide
{insert part #} Page 44
Please be sure the
transmitter is turned
on, the IR port is
pointed directly at the
transmitter, at a
distance less than 4 ft.
(1.2 m) and try again.
transmitter is turned
on, the IR port is
pointed directly at the
transmitter, at a
distance less than 4 ft.
(1.2 m) and try again.
Programmer infrared
port not aligned with
transmitter infrared
port.
Transmitter turned off.
Programmer batteries
are low.
Transmitter defective.
Programmer
defective.
port not aligned with
transmitter infrared
port.
Transmitter turned off.
Programmer batteries
are low.
Transmitter defective.
Programmer
defective.
Re-align ports and try
again.
Turn on transmitter.
Recharge programmer
batteries.
Contact Data Critical
Corp. regarding repair
and/or replacement of
transmitter or
programmer.
again.
Turn on transmitter.
Recharge programmer
batteries.
Contact Data Critical
Corp. regarding repair
and/or replacement of
transmitter or
programmer.
Programming Failed
Programmer was
unable to send
programming changes
to transmitter.
Programmer was
unable to send
programming changes
to transmitter.
Interference or
ambient light noise.
Programmer batteries
are low.
Transmitter defective.
Programmer
defective.
ambient light noise.
Programmer batteries
are low.
Transmitter defective.
Programmer
defective.
Re-align ports and try
again.
Recharge programmer
batteries.
Contact Data Critical
Corp. regarding repair
and/or replacement of
transmitter or
programmer.
again.
Recharge programmer
batteries.
Contact Data Critical
Corp. regarding repair
and/or replacement of
transmitter or
programmer.
Unable to Access Port
Probably in Use
Probably in Use
Programmer error.
Reboot programmer
by inserting stylus in
the reset button
located on the back on
the programmer.
Contact Data Critical
Corp. regarding repair
and/or replacement of
programmer.
by inserting stylus in
the reset button
located on the back on
the programmer.
Contact Data Critical
Corp. regarding repair
and/or replacement of
programmer.
You must select or get
a bed number and one
or more pagers
a bed number and one
or more pagers
Programmer SEND
button was pressed
before selecting a bed
number or pager
numbers.
button was pressed
before selecting a bed
number or pager
numbers.
Follow procedure to
select bed and/or
pager numbers and
tap SEND again.
select bed and/or
pager numbers and
tap SEND again.