Given Imaging Limited CAPDH-2 Manual Do Utilizador

Página de 250
PillCam Capsule Endoscopy
80
PillCam Recorder—Capsule Pairing (DR3 only)
PillCam Recorder—Capsule Pairing (DR3 only)
PillCam recorder DR3 operates only with capsules that were paired to it. This allows the PillCam recorder 
to recognize images from the correct capsule. Pairing is initiated when the PillCam recorder recognizes 
transmission from a designated capsule. From that point on, the recorder is associated with the capsule. 
The recorder and capsule are then paired.
Designation of the capsule to be used in the procedure may be performed during patient check-in or 
immediately before capsule ingestion. The following section describes these two scenarios.
Capsule Designation During Patient Check-in
While entering the patient and procedure data during check-in, type the capsule ID in the Capsule ID field 
or use a barcode scanner. The capsule ID is displayed on the capsule package. 
During recorder initiation, the unique capsule ID will be transferred into the PillCam recorder together with 
the rest of the initiation data. It will appear on the PillCam recorder LCD together with the rest of the 
patient data.
As long as no transmission from the checked in capsule is received, pairing is not done. The 
Ready for 
pairing
 icon 
 remains in the top right corner of the screen.
If there are PillCam capsule transmissions received by the PillCam recorder, which are not from the 
designated capsule, the capsule IDs of these other capsules (only of the checked-in capsule type) are 
temporarily displayed on the PillCam recorder screen. 
Ready for pairing
 icon 
 remains in the top right 
corner of the screen.
When you open the designated capsule box, just before ingestion, the capsule starts blinking and 
transmitting, each capsule type in its unique way. The PillCam recorder will automatically pair to the 
designated capsule and start recording when it receives transmissions from it.
The 
Pairing success
 icon 
 will appear in the top right corner of the screen.
!
Warning
The PillCam recorder starts recording only after it is paired with the 
capsule
. If pairing is not complete, no video can be produced. Make sure 
you see the Pairing Success icon 
 
on the screen and the capsule LED 
on top of the recorder blinking in BLUE 
 before capsule ingestion.
Capsule ID