Agfa Gevaert N V 5364A Manual Do Utilizador

Página de 266
147
2900H EN 20071108
Advanced operation (Key-operator mode)
D
RYSTAR
 5500/5503 R
EFERENCE
 
MANUAL
2
To determine the reference values for geometry, measure the distances A and B 
of the geometric square on the test image.
3
Record these values as reference dimensions A
ref
 
and
 B
ref
 
on Chart 4 
(‘Drystar 5500/5503 Geometric Consistency Control Chart’). Refer to 
.
These charts will be used for the annual quality test. For more information, refer to 
 on page
4
Save this film for future reference.
Make sure to measure distance A from the left edge of the left line to the 
right edge of the right line and distance B from the upper edge of the 
upper line to the lower edge of the lower line.
We strongly recommend using a 30 cm (12 inch) machinist scale with 
0.5 mm divisions (1/64 inch).