Drystar 5300 Imager: Determination of Operating Levels 
Chart 1
Imager Type: __________
Serial #: _________________
Date ____________________
Film Type: ____________
Emulsion #: ______________
Densitometer (Internal or External ) 
If External: Type ____________  
Serial #: _________________
Day 1
Day 2
Day 3
Day 4
Day 5
Month
Day
Initials
Base + Fog 
Average of 5 Values = operating (aim) level  Base + Fog
Mid Density  
Average of 5 Values = operating (aim) level "Mid Density"
Dens. Difference 
Average of 5 Values = operating (aim) level  Density Difference 
Dmax
Average of 5 Values = operating (aim) level  Dmax
Step 1: Print QC Test images on five consecutive days. Record the optical densities 
measurements in the tables below. After five days, average the values to determine 
the operating (aim) levels for each of the parameters.
 
Step 2: Copy the operating (aim) levels to Charts 2A/B (’Daily Density Control Chart’)