5.3.2 Pre-acquisition commands
¾ Press the Start Calibration button to start calibrating the processing parameters
used on the acquired data.
¾ Push the trolley a few meters then press the End Calibration button to save the
parameters.
¾ Double-click the gray bar under the Menu bar (shown) to access the advanced
settings menu. See section 5.3.3 Advanced Setting, for more information.
Gain calibration
!
NOTE
Calibration can be repeated as often as required during a survey. It is most
useful whenever the ground conditions change, i.e. from asphalt to concrete, or
from concrete to grass.
useful whenever the ground conditions change, i.e. from asphalt to concrete, or
from concrete to grass.
¾ Press the Skip Calibration button to avoid performing the gain calculations. If you
skip calibration, you can only view pre-existing radar maps.
¾ The indicator lights represent the radar (RADAR), battery (BATT), acquisition
speed (SPEED) and acquired sweeps (SWEEP) show the current activity status of
the system:
the system:
.
¾ Press a color scale button to select the desired palette for viewing radar maps:
,
,
,
,
,
,
,
,
¾ Zoom in on the maps up by selecting one of these buttons:
¾ Press the End Program button to shut down the software.
ACTIVE
NOT ACTIVE
Gray Bar