Each QRS complex is evaluated as follows:
1.
Select the colored profile from the Patient Screening Tool that best matches the amplitude of the QRS (Figure 5). For
biphasic signals, the larger peak should be used to determine the appropriate colored profile. The QRS peak must fall
within the window bounded by the dotted line and the peak of the colored profile.
Note: ECG gains > 20 mm/mV are not permitted. If, when printed at the maximum 20 mm/mV gain, the QRS
Note: ECG gains > 20 mm/mV are not permitted. If, when printed at the maximum 20 mm/mV gain, the QRS
peak does not reach the minimum boundary (dotted line) of the smallest colored profile, that QRS complex is
deemed unacceptable.
Figure 5:
2. SELECT the colored profile. The largest
QRS peak must be within a Peak Zone.
Selecting the colored profile
2.
Align the left edge of the selected colored profile with the onset of the QRS complex. The horizontal line on the
colored profile should be used as a guide for isoelectric baseline alignment.
colored profile should be used as a guide for isoelectric baseline alignment.