Orthosoft Inc. R20S21A Manual Do Utilizador
SmartKnee Surgical Technique
31 of 58
2. Insert a USB key (must be FAT32 and NTFS). Note that Password protected USB keys are not
supported by the system controller.
3.
Click on the “System Utilities” tab and then click on the “Copy Logs” button
. The
“Copy Logs” button
will not be available until a valid USB key is connected to the system
4. The system will copy the logs on the USB key and will inform the user when the archiving is
successful.
4.4 How to activate the Field Issue recorder function
The Field Issue Recorder is a utility that extracts useful data for the investigation of anomalies. It transfers a
file containing necessary system logs to a removable USB drive. This file can later be sent to a customer
service representative for further investigations of the field issue.
file containing necessary system logs to a removable USB drive. This file can later be sent to a customer
service representative for further investigations of the field issue.
The Field Issue Recorder program is located in the "Maintenance" mode. To access the maintenance mode
please follow these steps:
please follow these steps:
1. Press the "System Utilities" tab
2. Click on the "System" button
3. Press "Ok" on the pop-up "Are you sure you want to switch to maintenance mode?"
4. In the "Maintenance" mode, launch the Field Issue Recorder program and follow the instructions.
2. Click on the "System" button
3. Press "Ok" on the pop-up "Are you sure you want to switch to maintenance mode?"
4. In the "Maintenance" mode, launch the Field Issue Recorder program and follow the instructions.
4.5 How to record snapshots during a surgical procedure
The Snapshot button can be pressed at any time during the procedure on the System Controller screen. All
images will be saved in the snapshot directory found in the surgery folder of the Copy Logs option.
images will be saved in the snapshot directory found in the surgery folder of the Copy Logs option.
4.6 How to shut down the System Controller
To shut down the system controller, follow the below steps:
1. Exit current procedure, if any.
2. Click on the “System Utilities” and then on the “shutdown” button.