Instrument 
Manuf. No 
Qty  Sterilization 
and specific 
cleaning 
instructions 
Additional notes 
Smart Tools Knee Pod Kit 
20-8011-049-00 
ETO 
 
Single use  
 
Disposable 
 
  
38 mm screws 
 
 
 
 
3.5mm Hex head screw x 38mm
 
20-8000-000-18 
Autoclave 
 
Provided Sterile 
 
Single use 
75 mm screws 
 
 
 
 
3.2mm Headless Trocar drill pin   x 1      (a) 
20-8000-000-15 
Autoclave 
 
Provided Sterile  
 
Single Use 
Headless trocar drill pin, 75mm             (a) 
00-5901-020-00 
Autoclave 
 
Provided Sterile  
 
Single Use 
 
(a)  These three pins are equivalent in material and dimensions. The 
20-8000-000-16 is provided non-sterile. 
 
 
Do not use pins or any other fasteners than those recommended above
 
 
.
 
5.1.6  POD  specific information 
Note that the Pod’s consumption is optimized to provide two-hours of use from power-up, which 
is when the button is pushed on the Pod’s side. 
These Pods are contained in a custom plastic tray preventing activation prior to use including 
during ethylene oxide sterilization. 
Pods complies with IPx7 standard to be protected against the effects of temporary immersion in water
 
 
Pods must be kept in their plastic tray during ethylene oxide sterilization to prevent potential 
explosions. 
 
Pods should be properly connected to the instrument to provide strong fixation between both 
parts. 
 
Pods should be connected to the instrument according to the same symbol visible on the Pod 
membrane and on the instrument