Orthosoft Inc. R20S21A Manual Do Utilizador
SmartKnee Surgical Technique
41 of 58
Instrument
Manuf. No
Qty Sterilization
and specific
cleaning
instructions
cleaning
instructions
Additional notes
Smart Tools Knee Pod Kit
20-8011-049-00
1
ETO
Single use
Disposable
38 mm screws
3.5mm Hex head screw x 38mm
20-8000-000-18
4
Autoclave
Provided Sterile
Single use
75 mm screws
3.2mm Headless Trocar drill pin x 1 (a)
20-8000-000-15
3
Autoclave
Provided Sterile
Single Use
Headless trocar drill pin, 75mm (a)
00-5901-020-00
3
Autoclave
Provided Sterile
Single Use
(a) These three pins are equivalent in material and dimensions. The
20-8000-000-16 is provided non-sterile.
Do not use pins or any other fasteners than those recommended above
.
5.1.6 POD specific information
Note that the Pod’s consumption is optimized to provide two-hours of use from power-up, which
is when the button is pushed on the Pod’s side.
is when the button is pushed on the Pod’s side.
These Pods are contained in a custom plastic tray preventing activation prior to use including
during ethylene oxide sterilization.
during ethylene oxide sterilization.
Pods complies with IPx7 standard to be protected against the effects of temporary immersion in water
Pods must be kept in their plastic tray during ethylene oxide sterilization to prevent potential
explosions.
explosions.
Pods should be properly connected to the instrument to provide strong fixation between both
parts.
parts.
Pods should be connected to the instrument according to the same symbol visible on the Pod
membrane and on the instrument
membrane and on the instrument