Orthosoft Inc. R20S21A Manual Do Utilizador
SmartKnee Surgical Technique
47 of 58
5.2 Transporting and Storing
5.2.1 System Controller
System Provider is coming from the kit [KT-8011-010-01]
The System Controller should be placed back in its Carrying Case. Make sure to securely bundle the chord
in the case and secure with the Velcro strap.
in the case and secure with the Velcro strap.
If the System Controller needs to be shipped with a carrier (ex. FedEx), place the Carrying Case in the
Shipping Case. The Shipping Case is specifically designed to withstand carrier shipping conditions. Please
contact Customer Service for more information.
Shipping Case. The Shipping Case is specifically designed to withstand carrier shipping conditions. Please
contact Customer Service for more information.
To avoid any damages on the cables, make sure they are not bent at sharp angles
If wires are not properly packaged, they might interfere the system controller touch screen
and damage the system.
and damage the system.
Make sure that no objects are leaned on the system controller touch screen when closing
the carrying case. Any objects on the touch screen may damage it during the system
transportation or storage.
the carrying case. Any objects on the touch screen may damage it during the system
transportation or storage.
Make sure the computer is properly shutdown before closing the carrying case. If not, the
system may overheat and be damaged.
system may overheat and be damaged.
Do not leave the system in an uncontrolled environment where the storage temperature is
below -20 C or above 70°C. This may damage the equipment.
below -20 C or above 70°C. This may damage the equipment.
The System Controller must never be submerged in water or exposed to high level of
humidity.
humidity.
Use of transducers and cables other than those specified, with the exception of transducers
and cables sold by Zimmer as replacement parts for internal components, may result in
increased emissions or decreased immunity of the System Controller.
and cables sold by Zimmer as replacement parts for internal components, may result in
increased emissions or decreased immunity of the System Controller.
The System Controller should not be used adjacent to or stacked with other equipment. if
adjacent or stacked use is necessary, the System controller should be observed to verify
normal operation in the configuration in which it will be used.
adjacent or stacked use is necessary, the System controller should be observed to verify
normal operation in the configuration in which it will be used.
5.2.1 Smart Tools Knee Pod Kit
Please contact Customer Service for more information about transportation and storage of Pods