ActiGraph LLC 01 Manual Do Utilizador

Página de 15
   
 
 
 
www.theactigraph.com
 
P a g e
 | 14
 
49 E. Chase St.             Pensacola, FL 32502              tel 850.332-7900             fax 850-332-7904 
 
Device Connected to PC
 
Red LED (Fault Indicator) 
2 Flashes 
Li-Ion Battery is Faulty 
3 Flashes 
A hardware failure occurred while recording data.   
Contact customer support at 
Green LED  
1 Flash 
Battery Charging
 
Multiple Flashes 
Communicating with PC via USB 
Steady On 
Battery Fully Charged 
Device Not Connected to PC
 
Red LED (Fault Indicator) 
No Flashing 
(LED Off) 
Normal operating condition or battery dead 
2 Flashes 
Low Battery (use ActiLife Lifestyle software to check for remaining battery life). The 
unit needs to be recharged. 
3 Flashes 
-  Unexpected Battery Failure (Temporary battery power loss) 
                   or 
-  Battery Level has fallen below 3.1V and the unit has entered Halt Mode 
Green LED
 
No Flashing 
(LED Off) 
Actively collecting data (“Flash Mode” disabled) or battery dead 
1 Flash 
-  Delay before start mode (the LED always flashes prior to starting data collection) 
Actively taking data ("Flash Mode" enabled – not recommended)
 
2 Flashes 
N/A
 
3 Flashes 
-  End of memory reached (Device no longer collecting data) 
Battery  died while unit was in delay before start mode (no data collected on 
device)
 
Wireless Feature Enabled
 
Yellow LED (Wireless Indicator)
 
1 Flash Every 5 Seconds 
-  Actively paired with and receiving Heart Rate data 
1 Flash Every Second 
-  Actively paired and communicating with ActiLife or ActiLife Mobile 
Note:  The Red LED will ALWAYS flash to indicate LOW BATTERY regardless of whether "Flash Mode" is enabled or 
disabled.  If a “stop time” (optional) has been reached, the Green LED will stop flashing all together regardless of its 
previous state.
 
Table 7 – wGT3X+ and wActiSleep+ LED Reference Table