ActiGraph LLC 03 Manual Do Utilizador

Página de 21
 
 
 
 
actigraphcorp.com
 
49 E. Chase Street Pensacola, FL 32502   
tel (850) 332-7900 
fax (850) 332-7904 
TAS14DOC07 User’s Manual 
                    Page 14 of 21
 
Stopping the device 
In order to stop the device from collecting data, the device needs to be plugged into a computer with 
the ActiLife Software running.  One easy way to stop the device from collecting data is to set a stop time 
during initialization.  Once a device is told to stop, it will go into HALT mode and go into a low power 
stopped state.  Another way to get a device to stop collecting data is to plug it into a computer with the 
ActiLife Software running and update the firmware on the device.  This also will put it into HALT mode 
and have the device go into a lower power stopped state. 
Unexpected operations and events 
If a device performs in an unexpected manner and does not collect data, does not perform as long as 
expected, will not download data, or acts in any other manner that is unexpected, notify ActiGraph 
immediately.  ActiGraph’s Customer Service can be contacted at the phone number (877) 497-6996 and 
is available every week, Monday through Friday, 8A.M. to 5P.M Central Standard Time.  Customer 
Service may also be contacted through email at 
.      
Household Interference 
Within a household environment many different factors can negatively affect the device.  Putting a 
device close to hot sources, such as a fireplace and baseboard heater, will melt the housing of a device 
and destroy it.  Putting a device in the freezer for extended periods of time will reduce the capacity of 
the battery’s lifespan and negatively affect its performance.  Putting a device in a microwave will destroy 
various parts of the device and possibly completely destroy it.  Wireless internet connections, wireless 
television service, and other wireless services will not negatively affect the performance of the device.  
As a rule of thumb, do not put the device in extreme cold or hot temperatures and do not apply any 
electrical sources to the device unless explicitly approved and described by ActiGraph. 
Battery 
All ActiGraph devices use a lithium polymer rechargeable battery that has a maximum voltage of 
approximately 4.20 volts and have a total life expectancy of five (5) years.  Even though these batteries 
are replaceable, only ActiGraph personnel are to replace the batteries.  At 3.1 volts the devices enter a 
low voltage mode state (HALT mode). 
In order to slow the aging of the lithium polymer battery, it is recommended that devices be stored in a 
partially charged state (40 – 60% battery capacity) in low ambient temperatures.  Every three (3) months 
plug the devices in to monitor their voltages and to extend the life expectancy of the battery.  Table 5 
demonstrates the impact of storage temperature and charge state on battery long term capacity.