Wave Gain Tab (Patient)
Patient Setup Functions
5 - 54
0070-10-0690-01
Panorama™ Operating Instructions
5.6.3
Sidebar Buttons
The following sidebar buttons are used to expand the functionality of this tab.
Restore Previous Settings
Select the Restore Previous Settings sidebar button to return patient display settings to 
those in use when the tab was opened.
Normal Screen
Select the Normal Screen sidebar button to return the display to the full screen mode, and 
close all tabs, message boxes, and dialog boxes.
5.6.4
Troubleshooting
This section lists some of the potential messages and issues that may occur while using the 
patient Wave Gain tab.
MESSAGE/ISSUE *
REASON
SOLUTION
No Patient Selected! 
This message is displayed in 
the demographic field when 
either a patient tile has not 
been selected or an empty tile 
was selected.
Select an active patient tile.
The patient’s wave gain 
settings did not change
The Done button was not 
selected after the patient’s 
wave gain settings were 
changed.
Adjust the patient’s wave gain 
settings and select the Done 
button when the settings are 
completed.
*
Messages are shown in all bold text.