Using the LifeVest
Battery status indicators
Screen shows
What it means
What to do
There is no battery in the
charger
Insert the battery into the charger. Leave one battery in the
charger while you use the other battery.
Battery is charging
Let the battery charge. This can take up to 4 hours.
Battery is fully charged,
ready for use in monitor
Leave battery in the charger until ready to exchange
battery.
Battery is being tested
Let the test run, which can take up to 12 hours. Test verifies
battery function as part of normal routine. After test, battery
will be charged, which can take another 4 hours. So the
will be charged, which can take another 4 hours. So the
whole process can take up to 16 hours.
If you need to use the battery sooner, remove battery from
If you need to use the battery sooner, remove battery from
charger as soon as you see this on the screen. Then put the
battery back into the charger to skip the test and charge the
battery back into the charger to skip the test and charge the
battery. The test will be performed the next time the battery
is plugged into the charger.
Battery has a problem, but
still might hold a charge
The battery may charge, but will take longer than normal.
Battery can be used, but should be replaced.
Call ZOLL Lifecor.
Battery can be used, but should be replaced.
Call ZOLL Lifecor.
Charger has a problem, and
cannot charge the battery.
Do not leave a battery in the charger. Battery is not being
charged.
Remove the battery and reinsert it firmly to try again.
Call ZOLL Lifecor.
Remove the battery and reinsert it firmly to try again.
Call ZOLL Lifecor.