Using the LifeVest 
Periodically clean and inspect the system 
How to clean the garment 
Specific details about laundering the garment are in section 4. 
How to clean the non-washable items 
Unplug the battery charger and disconnect the power cord before cleaning. 
Non-washable items such as the battery, charger, cables, ECG electrodes, and 
therapy pads may be cleaned using a soft cloth sparingly dampened with 
Formula 409
®
 all-purpose cleaner or equivalent cleaning solution. 
When you clean these items, keep in mind: 
y
  Don’t apply liquids directly to any of the non-washable items, as they contain 
electronic components that can be damaged. 
y
  Don’t attempt to clean any electrical contacts or connectors. 
y
  Don’t use any cleaning solution on the garment. 
Inspection 
Inspect your system periodically. If you should notice any of the following 
conditions, please notify ZOLL Lifecor as soon as possible: 
y
  Cracks in the housing of the monitor, battery, or charger. 
y
  Cracks in the therapy pads. 
y
  Tears in the garment or holster. 
y
  Blue gel leaking from the therapy pads at any time other than when 
defibrillation is about to occur or has just occurred. 
1-800-543-3267 