Responding to alerts 
Check belt message 
After you get a number of belt problem screens, you may 
get this message. It means that the monitor is still not 
receiving a good signal from the electrode belt. 
If you get this message: 
1    Disconnect the electrode belt from the monitor. 
2    Take off the garment and make sure that nothing is 
covering the metal surfaces of the ECG electrodes. 
3    Put a dab of unscented hand 
lotion, skin cream, or moisturizer 
on the surface of each ECG 
electrode (the round ones, not on 
the therapy pads).  
 
 
4    Put the garment on and connect the electrode belt to 
the monitor. 
While you’re wearing the LifeVest, tap the help button 
 
for reminders about what to do. 
If problems continue, call ZOLL Lifecor. 
1-800-543-3267 
Page 5-11