Transportation environment: -40°C°C~55°C, Relative humidity: 10%~95% 
Storage environment: -40°C~55°C, Relative humidity: 10%~95% 
2.5.2 Technical parameter   
Name  Headset type ECG recorder   
Model  RN141 
Channel  Single channel 
Technical parameter  Dynamic input range: 
For  superimposed  ±300mV  DC  bias  voltage,  with  the  change  of  the 
differential mode voltage at the speed of 125mV/s, amplitude modulation 
of 6mV (when the gain is set to be 10mm/mV), the recorder should have 
the  ability  to  respond  and  display.  The  magnitude  of  the  time-varying 
output signal is equivalent to the input change of no more than 10%. 
Input impedance: 
For the 10HZ sine wave signal, the input impedance should be greater 
than 10
 
MΩ. This requirement should be met within the range of ±300mV 
DC bias voltage. 
Common mode rejection: 
For the sinusoidal signal under the network power supply frequency, the 
common mode rejection should be at least 60dB, for the signal under 2 
times  network  power  supply  frequency,  the  common  mode  rejection 
should be at least 45dB. 
Systematic noise: 
When  the  input  terminals  connect  each  patient's  electrode  in  series 
through a resistance and capacitance network in parallel including 51kΩ 
resistor and a 47nF capacitor, the internal noise equivalent to the input 
should not exceed 50μV (peak - valley value) within 10s at random. 
Gain accuracy: 
When the gain is set to be 10mm/mV, the test signal of the output signal 
equivalent to input, the maximum amplitude error is ± 10%.