Cardinal Health Commercial Technologies LLC FUSE001 Manual Do Utilizador

Página de 47
38
Error Messages
Situation
Error Message
Corrective Action
While loading a new  
medication or a refill of a 
medication, you load the 
wrong pod into the pod tray.
Oops. Please remove the pod 
from the tray.
Sorry, the pod is invalid, could not 
be read, or has been 
discontinued.
Remove the incorrect pod, tap 
I’ve removed the pod, and load 
the correct pod.
While taking a medication, 
you remove a pod when you 
were not prompted to do so.
Oops.  Something went wrong.
Make sure the X is loaded.
(Where X is the medication name.)
Place the pod back into the 
pod tray, and then tap I’ve 
put in the right pod
While taking a medication, 
you are prompted to take 
more medications.
Please take X more tablets. 
(Where X is the number of  
additional medications the  
device prompts you to take 
from the pod).
If you wish to purposely take 
less than the prescribed dosage, 
tap Skip this Medication.
If you removed the correct 
number of medications from 
the pod, tap I took the correct 
amount
, and then contact  
Cardinal Health Technical 
Support.
While taking a medication,  
you are prompted to 
place medications back 
into the pod.
It seems too many pills have been 
removed. Please put X tablets 
back.
(Where X is the number of  
medications the device prompts 
you to return to the pod).
This message is accompanied by an 
audible alarm.
If you lost a medication or if a 
medication is damaged, tap  
A pill was lost/damaged.
If you removed the correct 
number of medications from 
the pod, tap I took the correct 
amount
, and then contact 
Cardinal Health Technical  
Support.
While refilling or  
discontinuing a medication, 
you did not remove the pod 
when prompted but you did 
tap I’ve removed the pod.
Please remove the pod from 
the tray.
Remove the pod from 
the tray, and then tap  
I’ve removed the pod.