Respironics Inc. 1127941 Manual Do Utilizador

Página de 100
 
 
Instructions for Use | Therapy Modes and Controls 
 
 
 
Overview 
17 
 
3.  Therapy Modes and Controls 
Overview 
Trilogy Evo Universal therapy modes can be used for invasive and noninvasive ventilation with all circuit types, including mouthpiece ventilation.  
Breath Types 
Trilogy Evo Universal can deliver the following breath types: 
o
 
Mandatory: Ventilator-initiated, time-cycled 
o
 
Assist-Control: Patient-initiated, time-cycled 
o
 
Spontaneous: Patient-initiated, patient-cycled 
Triggering and Cycling 
Patient triggers 
Auto-Trak is a flow trigger with rules that make patient-triggering and cycling more comfortable for the patient. The system uses multiple 
algorithms to detect the start and end of the breath. Also, it automatically adjusts the trigger and cycle sensitivity to optimize synchronization 
between the patient and the ventilator.  
Sensitive Auto-Trak is a more sensitive version of Auto-Trak 
Flow trigger initiates a breath when the patient’s inspiratory effort creates a flow equal to or greater than the flow trigger sensitivity setting. A 
lower number is more sensitive. As inspiratory flow begins to decrease, the device cycles to expiration when the patient flow is less than the 
percentage of peak flow, based on the flow-cycle sensitivity setting. 
Ventilator trigger 
Time Trigger is time-based, defined by the Breath Rate setting. 
Flow patterns 
Ramp wave pattern: airflow starts high and decreases throughout inspiration of the breath. 
 
Square wave pattern
airflow is generally constant throughout inspiration of the breath. 
 
Therapy Modes Overview 
Control Modes: breaths are assist-control or mandatory.  
• 
A/C-PC: Assist control – assist-control and mandatory breaths with pressure control 
• 
A/C-VC: Assist control – assist control and mandatory breaths with volume control 
Spontaneous modes: the patient initiates all breaths.  
• 
CPAP: Continuous positive airway pressure 
• 
PSV: Pressure support ventilation