Instructions for Use | Device Setup 
 
 
 
Connecting a Circuit 
33 
 
 
Connecting a Circuit 
Be certain that any breathing system filter used with this device complies with ISO 23328-1 and ISO 23328-2. · To prevent patient or ventilator 
contamination, we recommend you use a Respironics-approved main flow bacteria filter (Part Number 342077) on the patient gas outlet port.  
Filters not approved by Respironics may degrade system performance.  Note: A leak device is mandatory during invasive ventilation or when using 
a circuit with a non-vented mask. 
After you connect the circuit, you may calibrate the circuit. See chapter 7, “Device Options, Calibration.” 
For low tidal volumes, see the “Therapy Modes chapter”, section titled “Low Tidal Volume Therapy.” 
 
Passive Single Limb Circuits 
 
 
Bacteria filter 
Tubing 
Exhalation port 
 
1.  Connect the bacteria filter (1) on the circuit to the Inspiratory Port.