Instructions for Use | Alarms and System Messages 
 
 
 
Alarms and System Messages 
46 
 
Requirements 
 
Device 
performance 
during active 
alarm 
The alarm is automatically resolved when the delivered pressure comes within 5 cmH
2
O of the target 
patient pressure during the expiratory phase. 
 
Device will continue to operate. 
 
H
IGH 
I
NSPIRATORY 
P
RESSURE
 
Priority 
High 
Why it occurs 
During the inspiratory phase, the delivered pressure exceeds the target patient pressure by 5 cmH
2
or more  
What to do 
• 
Check the patient for: 
o
 
Coughing or excessive secretions 
o
 
Bronchospasms 
o
 
Tracheotomy tube stability 
 
• 
Check the ventilator for: 
o
 
Kinked, pinched, or blocked tubing 
o
 
Blocked leak device or blocked exhalation device 
o
 
Secretions in the HME 
Requirements 
Therapy mode must be SIMV-VC or A/C-VC 
Device 
performance 
during active 
alarm 
The alarm is automatically resolved when the delivered pressure falls within 5 cmH
2
O of the target 
patient pressure during the inspiratory phase. 
 
The device will automatically cycle to the expiratory phase and continue to operate. 
  
E
XTERNAL 
F
LOW 
S
ENSOR 
F
AILED
 
Priority 
High 
Why it occurs 
A flow sensor offset or sensor failure is detected when the device is in standby or delivering therapy. 
What to do 
Ensure the sensor and cable are connected properly and are not damaged. Clean if needed. Replace if 
necessary. 
Requirements 
Active flow or dual limb circuit 
Device 
performance 
during active 
alarm 
The alarm remains active until the flow sensor is replaced or reconnected. If the alarm occurs when 
delivering therapy, the device will continue to operate. 
If the alarm occurs when the device is in standby, you cannot start therapy. 
 
E
XTERNAL 
F
LOW 
S
ENSOR 
C
ABLE 
D
ISCONNECTED
 
Priority 
High 
Why it occurs 
The external flow sensor cable becomes disconnected from the ventilator during active therapy. 
Active flow or dual limb circuit is selected and an external flow sensor cable is not connected. 
What to do 
Ensure the sensor and cable are connected properly and are not damaged. Replace if necessary. 
Requirements 
Active flow or dual limb circuit 
Device 
performance 
during active 
alarm 
Prohibits starting therapy if the external flow sensor cable is not connected to the ventilator when the 
active flow or dual limb circuit is selected 
 
E
XTERNAL 
F
LOW 
S
ENSOR 
N
OT 
C
ONNECTED
 
Priority 
High 
Why it occurs 
The external flow sensor becomes disconnected from the external flow sensor cable during active 
therapy. 
What to do 
Ensure the sensor and cable are connected properly and are not damaged. Replace if necessary. 
Requirements 
Active flow or dual limb circuit 
Device 
performance 
Device continues to provide therapy at the set breath rate. 
Device will not trigger on a patient initiated breath.