10.  Trilogy Evo Universal Accessories 
To prevent adverse performance, use Trilogy Evo Universal only with accessories intended for use with this device, including all patient interfaces, 
masks, circuits, exhalation ports, and carrying cases. For a list of accessories, see the Trilogy Evo Universal accessories guide at: 
You must ensure accessories and parts are compatible before you connect a patient to the device. If you are using a remote alarm or nurse call, 
ensure you test the accessory before starting ventilation. 
The accessories described in this chapter are those included in your device package. For all other accessories, see the accessory’s instructions.  
Power Accessories 
For instructions, see “Power Management.” 
Patient Monitors 
FiO
2
 sensor 
An FiO
2
 sensor is installed in your device. For instructions on calibrating the sensor, see “Trilogy Evo Universal Options, O
2
 Sensor Calibration” 
Intended Use: The Philips Respironics FiO
2
 Sensor is an oxygen sensor device that is used for measuring the fraction of inspired oxygen (oxygen 
concentration in a percentage) delivered from the oxygen source to the airway of a patient circuit (such as hose and mask). 
Storage and Handling 
• 
Avoid rough handling that would damage the sensor. 
• 
Avoid exposing sensor(s) to rapid changes in pressure. 
• 
Avoid puncturing or damaging sensor membrane(s). 
Accidental release measures: The oxygen sensors contain a strong basic solution encapsulated in a plastic housing. Under normal operating 
conditions, the solution (electrolyte) is never exposed. In case of a leak, please observe the following instructions: 
Personal precautions, protective equipment, and emergency procedures 
Use personal protective equipment. Avoid dust formation. Avoid breathing vapors, mist, or gas. Ensure adequate ventilation. Evacuate 
personnel to safe areas. Avoid breathing dust.  
Personal protective equipment 
Eye/face protection: Safety glasses with side-shields or googles conforming to appropriate government standards such as ANSI (US) or EN 
166(EU) 
Skin protection: Handle with nitrile gloves. Gloves must be inspected prior to use. Use proper glove removal technique (without touching 
glove’s outer surface) to avoid skin contact with this product. Dispose of contaminated gloves after use in accordance with applicable 
laws and good laboratory practices. Wash and dry hands. 
Respiratory and body protection: Wear respiratory protection and full protective clothing tested and approved under appropriate 
government standards such as ANSI (US) or CEN (EU). 
Environmental precautions 
Prevent further leakage or spillage if safe to do so. Do not let product enter drains. Discharge into the environment must be avoided. 
Methods and materials for containment and cleaning up in case of leaking or spilling 
Contain spillage. Neutralize spill with soda ash or lime. Carefully place material into clean dry container and cover. Flush spill area with 
water. Avoid creating dust. 
Disposal