11.  Power Management 
 
Power Sources 
Trilogy Evo Universal can operate on AC (wall outlet) or DC (battery) power from several sources listed in descending priority below. 
1.  AC Power 
2.  External 24- or 12-volt battery (DC) such as a vehicle battery – requires an external battery cable 
3.  Detachable battery (DC) 
4.  Internal battery (DC) 
Battery Power Indicator 
The battery power indicator appears in the status bar at 
the bottom of the window. The indicator shows the 
status of batteries in use. 
 
Battery operating time depends on the characteristics 
of the battery and usage of the device. The capacity of 
the battery shown on the Battery Power Indicator is only 
an estimate.  
 
 
 
You can also view battery status in the following locations:  
• 
The battery status Monitoring Window – see “Monitoring Window” in the “About Trilogy Evo Universal” chapter. 
• 
The device information window: In the menu bar, tap the Options icon. 
 In the Options window, tap Information
Battery Status Indicator 
To view the battery status indicator, tap the battery power indicator in the Status Bar:  
 
Battery status 
This indicator shows: 
Battery status 
Estimated battery time remaining (for internal and detachable 
batteries) 
Estimated remaining battery power shown as a percent of total 
capacity (for external battery) 
 
AC Power 
AC power has the highest priority. When AC power is present, Trilogy Evo Universal uses that power to run the device and to charge the detachable 
and internal batteries.  
To use AC power:  
1.  Plug the socket end of the AC power cord into Trilogy Evo Universal’s power inlet. 
2.  Plug the pronged end of the power cord into an electrical outlet that is not controlled by a wall switch. 
3.  Secure the cord to the device using the retention clip.  
4.  Verify Trilogy Evo Universal is using AC power. You should see the green LED light next to the power button. If you do not see this light, 
contact Philips customer service.  
 
Periodically inspect the power cord for damage or signs of wear. Discontinue use and replace if damaged.