6.2 Error Messages and Troubleshooting 
 
● 
Battery Error 
If the “ 
  ” symbol appears, the Analyzer will not function as the battery 
power is too low. Please charge the battery as soon as possible. 
 
● 
Sampling Error 
Please remove the inserted test strip when the “
” symbol appears. This 
error could be caused by the following reasons: 
   
◆ 
The inserted test strip has been used or damaged. Please use a new test strip. 
    ◆  Do not use the test strips out of expiration date, the date is printed on the
 
foil 
wrapper
.
 Otherwise, you may get incorrect test results. 
   
◆ 
The test strip is dry or insufficient body fluid was sampled. Please use a new 
test strip and conduct a new test.   
   
The test strip is not Careomedic Fluorescent Immunoassay Test strip. 
Incompatible test strips may damage the Analyzer. 
 
● 
Temperature Error 
In order to get an accurate test result, please conduct tests between 5~40 °C 
(35~104 °F) . 
When the ambient temperature is too high or too low, the screen of the Analyzer 
will display the symbol blow. Error message “E01” will show on the screen when 
you press the button. Please move the Analyzer to an operating ambient 
temperature.   
 
When moving from a temperature that is outside the operating temperature range 
to an operating temperature, please wait for 10 minutes before conducting a test. 
 
● 
Motor Error