Innokas Yhtyma Oy VC150 Manual Do Utilizador

Página de 332
KO00065K
VC150 Vital Signs Monitor
9-5
Nellcor SpO
2
Description
CAUTIONS
Do not sterilize reusable sensors by irradiation, steam, or ethylene 
oxide. See the sensor manufacturer's instructions for cleaning, 
sterilization, or disinfecting methods.
Do not place SpO
2
 sensor on patient during magnetic resonance 
imaging (MRI). Adverse reactions include potential burns to patients as a 
result of contact with attachments heated by the MRI radio frequency 
pulse, potential degradation of the magnetic resonance image, and 
potential reduced accuracy of SpO
2
 measurements. Always remove 
oximetry devices and attachments from the MRI environment before 
scanning a patient.
Pulse oximetry readings and pulse signals can be affected by certain 
environmental conditions, sensor application errors, and certain patient 
conditions. See the appropriate sections of this manual for specific 
safety information.
Placing a sensor distal to an arterial line may interfere with adequate 
arterial pulsation and compromise the measurement of SpO
2
.
The sensor disconnect error message and associated alarm indicate 
that the sensor is either disconnected or the wiring is faulty. The user 
should check the sensor connection and, if necessary, replace the 
sensor, interconnect cable, or both.
Placing a sensor on a polished or an artificial nail may affect accuracy.
Patient safety:
Do not place any clip-on sensor in a patient’s mouth, on their nose or 
toes, on their thumb, or across a child’s foot or hand.
Prolonged monitoring or incorrect sensor application can cause skin 
irritation or impaired circulation. Observe the sensor site frequently to 
assure adequate distal circulation. Sensor sites should be checked at 
least every 2 hours and rotated at least every 4 hours. Refer to 
instructions supplied with sensor.
If the sensor is not applied properly, the patient’s skin could be injured or 
the ability of the monitor to measure oxygen saturation could be 
compromised. For example, a clip-on sensor should never be taped 
shut. Taping the sensor could damage the patient’s skin or impair the 
venous return, thus causing venous pulsation and inaccurate 
measurement of oxygen saturation.
Excessive pressure from the sensor may cause necrosis of the skin.
The operating range for respiration rate is 4 to 40 breaths per minute. 
Use on patients with respiration rates outside this range may result in 
inaccurate displayed respiration rate values.