Innokas Yhtyma Oy VC150 Manual Do Utilizador
KO00065K
VC150 Vital Signs Monitor
3-19
Getting started: Menu selection for setup screens
Menu selection for setup screens
1. Select Monitor Setup and subsequent tab sheets.
2. Select Audible & Visual, NIBP, SpO
2
or Temperature to make selections or to
modify the settings.
3. When you are finished with the configuration, select the home icon to return
to clinical mode.
Audible & Visual
An audible signal is used to indicate the presence of an alarm limit violation. If
audible signals need adjustment, these can be adjusted in the Monitor Setup >
Audible & Visual screen.
audible signals need adjustment, these can be adjusted in the Monitor Setup >
Audible & Visual screen.
The first tab in Monitor Setup is Audible & Visual. This allows you to adjust
Alarm Volume, Day Volume, Night Volume, Alarm Light Brightness, Day
Display Brightness and Night Display Brightness. The currently active day/
night mode is automatically determined by a light sensor on lower right part of
the monitor touch screen. The tab sheet displays the mode for both volume and
brightness above sound level bars. Sound levels can be lowered by selecting left
part of the sound level bar or raised by selecting the right part of the sound level
bar. These settings affect all applicable monitor screens. If the alarm sound level
cannot be set as low as required, it means that the minimum alarm sound level
has been set in the Monitor Setup > Advanced > Default Setup > General to be
something other than 0.
Alarm Volume, Day Volume, Night Volume, Alarm Light Brightness, Day
Display Brightness and Night Display Brightness. The currently active day/
night mode is automatically determined by a light sensor on lower right part of
the monitor touch screen. The tab sheet displays the mode for both volume and
brightness above sound level bars. Sound levels can be lowered by selecting left
part of the sound level bar or raised by selecting the right part of the sound level
bar. These settings affect all applicable monitor screens. If the alarm sound level
cannot be set as low as required, it means that the minimum alarm sound level
has been set in the Monitor Setup > Advanced > Default Setup > General to be
something other than 0.