Mirion Technologies Oy MBD-PD-2 Manual Do Utilizador

Página de 29
 
MBD-2 Operating Manual 
 
19 
15-00167, Rev 1 
c.  Remove the screws and set the cover aside on a clean surface. 
 
NOTE: the screws can be withdrawn without removal where the elastomer gasket material on 
the battery cover will hold the screw in place permitting removal, thus minimizing the risk of 
screw misplacement. 
 
 
Figure 18: Internal Battery Cover and Compartment
 
Internal Battery Cover  
 
Detector Battery CR1025, 2 each 
Torx (T6) screw 3each 
 
Real Time Clock (RTC) Battery, 1 each 
Main Battery Compartment 
 
Battery Orientation Markings 
 
 
d.  Replace two detector batteries and one RTC battery using Model CR1025, Renata 
preferred. Ensure battery orientation is Positive (+) side up.  The internal battery case 
features imprints for battery orientation and model number. 
 
 
 
e.  Dispose the used batteries in accordance with site requirements. 
 
Note:  the detector and real-time clock batteries shall be replaced together during scheduled 
maintenance.  
 
 
WARNING 
If the batteries appear to have leaked, i.e.; there is white residue on the battery 
and/or internal battery compartment surfaces; use personnel protective 
equipment (i.e.; gloves, safety glasses, etc.) as applicable.