DS1000Z Programming Guide
:TRIGger:RUNT:BLEVel
Syntax
:TRIGger:RUNT:BLEVel <level>
:TRIGger:RUNT:BLEVel?
:TRIGger:RUNT:BLEVel?
Description
Set or query the trigger level lower
limit in runt trigger. The unit is the same as the current
amplitude unit.
Parameter
Name
Type Range
Default
<level>
Real
(± 5 × VerticalScale from the screen center) -OFFSet 0
Explanation
. For OFFSet, refer
Return
Format
The query returns the trigger level lower limit in scientific notation.
Example
:TRIGger:RUNT:BLEVel 0.16 /*Set the trigger level lower limit to 160mV*/
:TRIGger:RUNT:BLEVel? /*The query returns 1.600000e-01*/
Related
Command