Rigol MSO1074Z-S 4-channel oscilloscope, Digital Storage oscilloscope, MSO1074Z-S Manual Do Utilizador
Códigos do produto
MSO1074Z-S
RIGOL
DS1000Z Programming Guide
2-43
:MATH:OPTion:SENSitivity
Syntax
:MATH:OPTion:SENSitivity <sens>
:MATH:OPTion:SENSitivity?
:MATH:OPTion:SENSitivity?
Description
Set or query the sensitivity of the logic operation. The default unit is div (namle the
current vertical scale).
Parameter
Name
Type
Range
Default
<sens>
Real
0 to 0.96div, the step is 0.08div
0
Explanation
This command is only applicable to the logic operations (A&&B, A||B, A^B and !A).
Return
Format
The query returns the sensitivity in scientific notation.
Example
:MATH:OPTion:SENSitivity 0.24 /*Set the sensitivity to 0.24div*/
:MATH:OPTion:SENSitivity? /*The query returns 2.400000e-01*/
Related
Command
:MATH:OPTion:DIStance
Syntax
:MATH:OPTion:DIStance <dist>
:MATH:OPTion:DIStance?
:MATH:OPTion:DIStance?
Description
Set or query the smoothing window width of the differential operation (diff).
Parameter
Name
Type
Range
Default
<dist>
Integer
3 to 201
3
Explanation
This command is only applicable to the differential operation (diff).
Return
Format
The query returns an integer between 3 and 201.
Example
:MATH:OPTion:DIStance 25
Related
Command
:MATH:OPTion:ASCale
Syntax
:MATH:OPTion:ASCale <opt>
:MATH:OPTion:ASCale?
:MATH:OPTion:ASCale?
Description
Enable or disable the auto scale setting of the operation result or query the status of the
auto scale setting.
Parameter
Name
Type
Range
Default
<opt>
Bool
{{0|OFF}|{1|ON}}
0|OFF
Explanation
When the auto scale is enabled, the instrument will automatically set the vertical
scale to the best value according to the current operator, the vertical scale and the
horizontal timebase of the source.
Sending this command will modify the auto scale status of all operations.
Return
Format
The query returns 0 or 1.
Example
:MATH:OPTion:ASCale ON