Lecroy WAVEACE 2004 4-channel oscilloscope, Digital Storage oscilloscope, WaveAce 2004 Manual Do Utilizador

Códigos do produto
WaveAce 2004
Página de 89
WaveAce Remote Control 
 
48 
920836 Rev A  
MISCELLANEOUS - COMM_HEADER, CHDR 
Description 
The COMM_HEADER command controls the way the oscilloscope formats responses to queries. 
There are three response formats:  
 
LONG, responses contain the long form of the header : C1:VOLT_DIV 200E-3 V 
 
SHORT, responses contain the short form of the header : C1:VDIV 200E-3 V 
 
OFF, headers and units are omitted: 200E-3 
Unless you specify otherwise, the SHORT response format is used by default. 
This command does not affect the interpretation of messages sent to the oscilloscope. Headers can be 
sent in their long or short form regardless of the COMM_HEADER setting. 
Command Syntax 
Comm_HeaDeR <mode> 
<mode>:= {SHORT, LONG, OFF} 
Query Syntax 
Comm_HeaDeR? 
Response Format 
Comm_HeaDeR <mode> 
Example 
The following instruction sets the response header format to LONG: 
CHDR LONG