Lecroy WaveSurfer 3024 4-channel oscilloscope, Digital Storage oscilloscope, WaveSurfer 3024 Manual Do Utilizador

Códigos do produto
WaveSurfer 3024
Página de 305
C
H A P T E R  
O
N E
:
 
 Overview 
 
WM-RCM-E Rev D 
ISSUED: February 2005
 9 
N
N
O
O
T
T
E
E
:
: If  you use one of  the older trace labels, for example "TC", any 
response from the scope uses the new label; for example, it  substitutes 
F3 for TC. 
DATA  
Whenever a command or query uses additional data values, the values are expressed as ASCII characters. There 
is a single exception: the transfer of  waveforms with the command/query WAVEFORM, where the waveform 
can be expressed as a sequence of  binary data values. See Chapter 4, “Wavefor m Str ucture.” ASCII data can 
have the form of  character, numeric, string, or block data. 
CHARACTER DATA  
These are simple words or abbreviations to indicate a specific action.  
Example:  F3:TRA ON 
In this example, the data value ON commands the trace F3 to be turned on (the data value OFF will have the 
opposite effect). 
However, this can become more complex. In some commands, where you can specify as many as a dozen 
different parameters, or where not all the parameters are applicable at the same time, the format requires pairs 
of  data values. The first value names the parameter to be modified, while the second gives its value. Only those 
parameter pairs that are to be changed need to be indicated.  
Example: HARDCOPY_SETUP DEV,EPSON,PORT,GPIB 
In this example, two pairs of  parameters have been used. The first specifies the device as an EPSON (or 
compatible) printer, while the second indicates the GPIB port. While the command HARDCOPY_SETUP 
allows many more parameters, either they are not relevant for printers or they are left unchanged. 
NUMERIC DATA  
The numeric data type is used to enter quantitative information. Numbers can be entered as integers or 
fractions, or in exponential representation:  
F1:VPOS -5 Move the display of  Trace A downward by five divisions. 
C2:OFST 3.56 Set the DC offset of  Channel 2 to 3.56 V. 
TDIV 5.0E-6 Adjust the timebase to 5 µsec/div. 
Example: There are many ways of  setting the timebase of  the oscilloscope to 5 µsec/div: 
TDIV 5E-6 Exponential notation, without any suffix. 
TDIV 5 US Suffix multiplier U for 1E−6, with the (optional) suffix S for  
 seconds, or TDIV 5000 NS or TDIV 5000E-3 US