WaveSurfer 3000 Oscilloscopes
Use the
to change the scale factors of the trace.
Roll Mode
Roll mode displays, in real time, incoming points in single-shot acquisitions that appear to "roll" continuously
across the screen from right to left until a trigger event is detected and the acquisition is complete. The
parameters or math functions connected to each channel are updated every time the roll mode buffer is
updated, as if new data is available. This resets statistics on every step of Roll mode that is valid because of
new data.
Timebase must be set to 100 ms/div or slower to enable Roll mode selection. Roll mode samples at ≤ 5
MS/s.
NOTE: If the processing time is greater than the acquire time, the data in memory is overwritten. In this case,
the instrument issues the warning, "Channel data is not continuous in ROLL mode!!!" and rolling starts again.
RIS Sampling Mode
RIS (Random Interleaved Sampling) allows effective sampling rates higher than the maximum single-shot
sampling rate. It is used on repetitive waveforms with a stable trigger. The maximum effective RIS sampling
rate is achieved by making multiple single-shot acquisitions at maximum real-time sample rate. The bins
thus acquired are positioned approximately 20 ps (50 GS/s) apart. The process of acquiring these bins and
satisfying the time constraint is a random one. The relative time between ADC sampling instants and the
event trigger provides the necessary variation.
The instrument requires multiple triggers to complete an acquisition. The number depends on the sample
rate: the higher the sample rate, the more triggers are required. It then interleaves these segments (as
shown in the following illustration) to provide a waveform covering a time interval that is a multiple of the
maximum single-shot sampling rate. However, the real-time interval over which the instrument collects the
waveform data is much longer, and depends on the trigger rate and the amount of interleaving required.
Interleaving of sample in RIS sampling mode.